Interprets pacemaker and ICD interrogation data with programming optimization documentation. Use when reviewing device interrogations, managing pacemaker settings, or documenting ICD therapies.
Interprets pacemaker and ICD interrogation data with programming optimization documentation.
Over 1.2 million cardiac implantable electronic devices (CIEDs) are implanted annually worldwide. Device interrogation data contains critical information about arrhythmia events, lead function, battery status, and therapy delivery that directly impacts patient safety. Failure to recognize lead malfunction, inappropriate ICD shocks, or suboptimal pacing parameters can result in syncope, cardiac arrest, or unnecessary therapies that degrade quality of life.
The HRS (Heart Rhythm Society) Expert Consensus on CIED management mandates systematic interrogation review, remote monitoring enrollment, and structured follow-up. This skill ensures every device check is documented completely with actionable findings.
Battery Status Evaluation:
| Status | Definition | Action |
|---|---|---|
| BOL (beginning of life) | Full battery voltage | Routine follow-up |
| MOL (middle of life) | Adequate remaining battery | Routine follow-up |
| ERI (elective replacement indicator) | Battery nearing depletion | Schedule generator replacement within 3 months |
| EOL/EOS (end of life/service) | Critical battery depletion | Urgent generator replacement; device may revert to VVI |
Lead Parameter Thresholds:
| Parameter | Atrial Lead | RV Lead | LV Lead (CRT) |
|---|---|---|---|
| Pacing threshold | < 1.5 V @ 0.4 ms | < 1.5 V @ 0.4 ms | < 3.5 V @ 0.5 ms |
| Sensing (P/R wave) | > 1.5 mV | > 5.0 mV | N/A (unipolar) |
| Impedance | 200–1500 Ω | 200–1500 Ω | 200–1500 Ω |
Lead Malfunction Red Flags:
Pacing Mode Nomenclature (NBG Code):
Common Pacing Modes:
| Mode | Typical Indication |
|---|---|
| AAI(R) | Sinus node dysfunction with intact AV conduction |
| VVI(R) | Atrial fibrillation with slow ventricular response |
| DDD(R) | Sinus node dysfunction or AV block with sinus rhythm |
| DDI(R) | Paroxysmal AF with AV block (avoids tracking) |
Assess:
Episode Classification:
| Zone | Typical Rate (bpm) | Typical Therapy |
|---|---|---|
| VT Monitor | 130–170 | Monitor only, no therapy |
| VT-1 | 170–200 | ATP × 3, then cardioversion |
| VT-2/VF | > 200 | ATP during charging, then defibrillation |
| SVT discriminators | Variable | Withhold therapy if SVT confirmed |
For Each Stored Episode, Document:
Inappropriate Shock Causes:
Biventricular Pacing Assessment:
Optimization Parameters:
CRT Response Assessment:
Structured Interrogation Report Must Include:
Follow-Up Intervals (HRS Recommendations):